Determination of viscosity of polymer by using Ubbelohde viscometer

Abstract: viscosity experiment notes: 1) the temperature of constant temperature water bath has a great influence on viscosity measurement. The temperature fluctuation of water bath should be within the range of + 0.05 DEG C, and the temperature of the water in the constant temperature water bath should be uniform. 2) capillary Ubbelohde viscometer must use clean liquid soaked and washed with water, can be used for drying the rear 105. 3) must be fixed on the vertical viscometer in constant temperature water bath. Otherwise, it will affect the velocity of flow and cause the error of measurement result. 4) after each dilution, the liquid in the viscometer must be mixed evenly. The method is to clamp the hose of the C tube, receive the gas from the B tube, mix the liquid in the D ball of the viscometer evenly, and then rinse the liquid in the D ball.

1. Basic principles
1.1, the viscosity of polymer solutions has the following definitions:

(1): relative viscosity ηr representation as . Among them, η0, it’s the pure solvent viscosity. η is the viscosity of solutions at the same temperature. ηr is a dimensionless quantity.
For low shear rate polymer solutions, the values are generally greater than 1. Obviously, with the increase of the concentration of ηr solution will increase.
(2): specific viscosity: indicated by ηsp, is compared with the solvent viscosity increase the fraction.
(3): Specific viscosity
Ratio of solution to specific viscosity to solution concentration

                          











The viscosity of the solution when the solution is infinitely diluted is independent of the concentration, and its dimension is the reciprocal of the concentration cm3/g

1.2 One point method
2. Device diagram



















As a common three branch glass viscometer.It has a capillary with a diameter of R and a length of L. At the top of the capillary there is a small ball with a size of V. There are lines a and B on the top and bottom of the ball. In the experiment, the C pipe is closed when the liquid is pumped from the A tube to the B tube, and then the C tube is passed through the atmosphere so that the liquid level at the lower end of the capillary drops. The liquid in the capillary forms a gas suspension column and flows down the pipe wall at the lower end of the capillary. This can avoid the exit turbulence as possible, and the equivalent average (which is the liquid column to suspension column height) with A tube liquid level.
Following the principle of determination of the viscosity of the liquid to be tested: adding self A tube, B tube by liquid suction to the a line above the B pipe through the atmosphere, let it flow down, record level through a and B line t.

3, Operation procedures and matters attention
Take the measured liquid along the inner wall of the tube into the A viscometer ball D. The viscometer is vertically fixed in a constant temperature water bath. Water temperature should be 25 - 0.05 DEG C, the water level higher than the buffer (upper B tube ball ball), placed 15 minutes, B tube and C tube are respectively connected with a latex tube, C tube hose clamp, B pipe mouth pumping; the test solution level increased to buffer the ball slowly part of the first open B nozzle, and then release the C nozzle, the test solution in the tube falls naturally, with a stopwatch accurate record level self-testing a down to fixed line by line B at the outflow time, were determined for two times, two times the difference should not exceed 0.1 seconds, averaged two times for the test solution the outflow time (T).
Matters needing attention
1) the temperature of the constant temperature water bath has a great influence on the viscosity measurement. The temperature fluctuation of the water bath should be within the range of + 0.05, and the temperature of the water in the constant temperature water bath should be uniform.
2) capillary Ubbelohde viscometer must use clean liquid soaked and washed with water, can be used for drying the rear 105.
3) must be fixed on the vertical viscometer in constant temperature water bath. Otherwise, it will affect the velocity of flow and cause the error of measurement result.
4) after each dilution, the liquid in the viscometer must be mixed evenly. The method is clamped C pipe hose, B tube gas, the mixed liquid D in uniform ball viscometer, ball, and then pumped D buffer liquid in the washing ball viscometer determination of ball (A and B line between the ball on the B tube), capillary 2-3 times.