Determination Intrinsic Viscosity of ACR

Determination Intrinsic Viscosity of ACR
First, equipment and auxiliary equipment

① Φ0.56mm a Ukrainian viscometer

② a set of automatic viscometer (25 ℃ ± 0.01 ℃)

③ 3 # sand core funnel 3 (covered)

④ conical flask 3

⑤ 75ml volumetric flask 3

Second, reagents: analytical grade (chloroform), content ≥ 99.0%
Third, the intrinsic viscosity of the sample test: 1, configuration solution 2, measured viscosity

1, configure the solution

① measuring bottle logo: the required capacity of 75ml three bottles (should be kept dry) clearly identified by the number.

Note: High viscosity is used separately from the volumetric flask used for low viscosity samples.

② balance calibration: Calibrate the balance, adjust the level (adjust the bottom of the balance knob so bubbles in the middle of the circle), with 200 grams weight calibration;

When adjusting the balance, or weighing, you should close the doors, windows, etc., at the same time shut down the balance door to prevent the wind from affecting the balance. When weighing, the balance can be read after the number of stable, prohibit the balance of data beating larger than reading.

 '¢ with liquid: open in Wang automatic dosing device DP25 and the DP25 and electronic scales connected to the sample bottle into the electronic balance and add 0.075g ± 0.01 ɡ expected in the automatic dial-up device DP25 touch screen settings The required concentration of 0.003g / ml, click the dosing button to complete dosing.

④ Sample Dissolution: Place the sample bottle containing the concentration solution in P12, and set the sample-dissolving time. The sample is dissolved automatically by P12.

2, Viscosity Testing:

① open the viscosity of the instrument set sink temperature 25 ℃ (Chung Mong Intelligent Viscosity Measurement System IVS300 series temperature control accuracy of ± 0.01 ℃), to be stable after the temperature field into the chloroform solvent with Ubbelohde viscometer (with basket) Click the start test, the system will automatically complete the continuous test and remove the unqualified data, get T0; the same way to measure the sample solution Flow time t.

② measure the t value of other bottles: the same way to measure the viscosity of the remaining two samples (tools can not be mixed).

③ by the viscosity of the software obtained in the sample viscosity of the sample solution.

Fourth, tool cleaning:

① Ubbelohde viscometer after use, the amount of chloroform 50ml beaker to take 25ml of chloroform added to the viscometer, use the ear wash up, then blow down, repeated three times (Note viscometer is not allowed to use water scouring, if water enters Should be dry), Ubbelohde viscometer upside down in the folder after the clip on the butterfly.

② volumetric flask and sand funnel were washed, with a 50ml beaker to take 25ml of chloroform into the volumetric flask, and then repeatedly shaken several times; sand funnel also 25ml chloroform wash shampoo, then wash the ear ball from the mouth Blowing out the waste a few times; Erlenmeyer flask with chloroform.

Note: All waste liquid into the waste collection bottle, warehousing unified treatment, after cleaning, measuring flask, sand funnel inverted placed in the tool plate, rack.