Zonwon was built on 2009 in Hangzhou,China. Aiming to provide "Auto Lab" for all the research department. We provide sample preparation and measuring instruments,the product line including: IVS Series Ubbelohde viscometer, D25/DP25 Pipette, A6 Furnace, P12 sample dissolving instrument and ,FHJ-8 Tube heater reactor. The best choice for high polymer material sample preparation and viscosity testing.
精选博文
How Temperature Affects Viscosity
How Temperature Affects Viscosity: Viscosity can be defined as the internal friction that occurs as a material flow, relative to itself....
2017年2月23日星期四
How to test the viscometer with Capillary tube GBGB17931-2003
A . 3 Reagents
Remarks: ISO 1628-1 :1998 1B type
capillary viscometer (Φ= 0. 88 m m ) can make sure the
Solution of flow time from 80 s to 200 s.
A . 4. 3
Heating device: to control the temperature at(110± 10) ℃
oil bath or other heating(1 10 ±1 0) ℃ device.
A. 4. 4 Magnetic stirrer or other
stirrer device.
A . 4. 5
Diameter (63-90) μm stainless
steel filter or micro-pores diameter for (40^-1 00)
μm glass filters
4. 8 Pipette: 25 m l. Automatic
liquid filling apparatus can be used.
A .4. 9 Erlenmeyer flask: 50 m l, 100 m l, with
glass stopper.
4. 10 Density
Meter: range (1.200^ 1.300) g/cm 3
A
. 5 Prepare Solvents and sample solution
If
the average time exceeds 1% of the initially data,this solvent should be
discarded and a new solvent should be prepared
A.
5. 2 Preparation of Sample solution
5.
2. 1 Weighed (0.1250±0.0010g) sample in 50 ML or 100 ML Cone bottle(a.4.9).
A.
5. 2. 2 Using Pipette (A.4.8) or Automatic pipetor measuring 25 ml phenol/
1.1.2.2 Tetrachloroethane Solvent(A.5.1) into cone bottle.
When
the sample dissolving finished, cool to room temperature.
A
. 6. 1 Preparation
Measuring
3 times of effluent under the circumstances of not emptying and reloading the
new solution. Make sure the max value and minimum values does not exceed
0.2s.Take average of 3 flow time as the sample solution flow time t.
Take
about 17 ml of solvent, using the same method as A . 6. 2 5 times , and make
sure the max and minimum values does not exceed 0.1s. Take average of 3 flow
time as the sample solution flow time t0. For the solvent in use measure the
average flow time a least one time a day.If the difference in the average
outflow time of two consecutive measurements is greater than 0.2s,the
viscometer should be cleaned..
Remarks:
Cleaning the viscometer:Put the viscometer in the solution
of sulfuric acid- potassium dichromate(i, 1 by volume) more than 12 h before
using or measured value is not parallel
washed with water first then washed with ionized wate (or Distilled water) at least 5 times, and then cleaned with acetone 1 time. At last dry the viscometer in less than 100 ℃.When the viscometer is in use,washed the test solution first,then washed with Trichloromethane, at last dry it.
I represent for viscosity of PET solution. using the
formula (A.1),In milliliters per gram (m L / g),
t0---Solvent
outflow time, in seconds (s)
Take the
average of the two parallel determinations as the test result. And fix the four
significant digits according G B / T 81 70- 1987.
Note: ISO
7792-1: 1997 provides that the viscosity number is one of the characteristic
properties of PET resin naming
Take two
parallel determination average as the test result and take three decimal places
according GB/T 8170-1987